Urinalysis (Dipstick and Microscopic): CLIA Requirements Mapped to 42 CFR 493

Urinalysis testing touches about eight CLIA standards, from test categorization (42 CFR 493.17) through QC (42 CFR 493.1256) and test reporting (42 CFR 493.1291). The table below maps each one to the Urinalysis records to keep ready.

Last reviewed Nov 24, 2026 · By the Sound Lab Consultants regulatory team · Sources listed below

Urinalysis requirement map

Use this map to find the regulation behind each Urinalysis requirement. Survey deficiency tags in CMS Appendix C are organized around these same sections.

Requirement42 CFR 493What to keep on file for Urinalysis
Test categorization (waived, moderate, high)42 CFR 493.17FDA categorization printout for your Urinalysis test system
QC42 CFR 493.1256Daily Urinalysis QC records, Levey-Jennings charts, corrective action log, control lot ranges
Calibration verification42 CFR 493.1255Six-month Urinalysis calibration verification and reportable range data with director review
Accuracy verification42 CFR 493.1236Twice-yearly Urinalysis accuracy checks (external PT, split samples or reference material) with acceptance criteria
Comparison of results across methods or sites42 CFR 493.1281Twice-yearly comparison if Urinalysis runs on more than one instrument, method or site
Method verification42 CFR 493.1253Urinalysis verification study: accuracy, precision, reportable range, reference interval check, director sign-off
Test report and critical values42 CFR 493.1291Critical value list, call log with read-back, report format showing Urinalysis reference intervals
PersonnelSubpart MTraining and competency records for everyone who runs Urinalysis

Proficiency testing

Not listed in Subpart I: no required PT, but accuracy must be verified at least twice a year (42 CFR 493.1236(c)).

Quick audit questions

Related tests

Full guide: Urinalysis (Dipstick and Microscopic) testing in the clinical lab: CLIA, PT and QC requirements on soundlabconsult.com, part of the Laboratory Test Compliance Library.

For laboratory compliance and operations only, not medical advice or patient care guidance.

Have a D-tag to answer?

Sound Lab Consultants maps each D-tag on your CMS-2567 to the regulation and Appendix C guidance and helps you build a correction that holds.

Book a free consultation

Related guides on soundlabconsult.com:

Frequently asked questions

Which CLIA regulations apply to Urinalysis?

Mainly 42 CFR 493.17 (categorization), Subpart H/I or 493.1236 (PT and accuracy), 493.1253 (verification), 493.1255 (calibration), 493.1256 (QC), 493.1281 (comparisons) and 493.1291 (reporting).

Does this page list D-tag numbers?

No. D-tags come from CMS Appendix C; this map points to the 42 CFR 493 sections those tags are based on so you can read the rule itself.

What does Urinalysis PT require?

Not listed in Subpart I: no required PT, but accuracy must be verified at least twice a year (42 CFR 493.1236(c)).

Sources