Urine Drug Screen (Drugs of Abuse): CLIA Requirements Mapped to 42 CFR 493
Urine Drug Screen testing touches about eight CLIA standards, from test categorization (42 CFR 493.17) through QC (42 CFR 493.1256) and test reporting (42 CFR 493.1291). The table below maps each one to the Urine Drug Screen records to keep ready.
Urine Drug Screen requirement map
Use this map to find the regulation behind each Urine Drug Screen requirement. Survey deficiency tags in CMS Appendix C are organized around these same sections.
| Requirement | 42 CFR 493 | What to keep on file for Urine Drug Screen |
|---|---|---|
| Test categorization (waived, moderate, high) | 42 CFR 493.17 | FDA categorization printout for your Urine Drug Screen test system |
| QC | 42 CFR 493.1256 | Daily Urine Drug Screen QC records, Levey-Jennings charts, corrective action log, control lot ranges |
| Calibration verification | 42 CFR 493.1255 | Six-month Urine Drug Screen calibration verification and reportable range data with director review |
| Accuracy verification | 42 CFR 493.1236 | Twice-yearly Urine Drug Screen accuracy checks (external PT, split samples or reference material) with acceptance criteria |
| Comparison of results across methods or sites | 42 CFR 493.1281 | Twice-yearly comparison if Urine Drug Screen runs on more than one instrument, method or site |
| Method verification | 42 CFR 493.1253 | Urine Drug Screen verification study: accuracy, precision, reportable range, reference interval check, director sign-off |
| Test report and critical values | 42 CFR 493.1291 | Critical value list, call log with read-back, report format showing Urine Drug Screen reference intervals |
| Personnel | Subpart M | Training and competency records for everyone who runs Urine Drug Screen |
Proficiency testing
Not listed in Subpart I: no required PT, but accuracy must be verified at least twice a year (42 CFR 493.1236(c)).
Quick audit questions
- Can you show Urine Drug Screen QC for every day patients were tested in the last month?
- When was the last Urine Drug Screen calibration verification, and did it cover low, mid and high values?
- Is the Urine Drug Screen verification study signed by the director and dated before the first patient result?
- Are Urine Drug Screen critical value calls logged with read-back?
Related tests
Full guide: Urine Drug Screen (Drugs of Abuse) testing in the clinical lab: CLIA, PT and QC requirements on soundlabconsult.com, part of the Laboratory Test Compliance Library.
Have a D-tag to answer?
Sound Lab Consultants maps each D-tag on your CMS-2567 to the regulation and Appendix C guidance and helps you build a correction that holds.
Related guides on soundlabconsult.com:
Frequently asked questions
Which CLIA regulations apply to Urine Drug Screen?
Mainly 42 CFR 493.17 (categorization), Subpart H/I or 493.1236 (PT and accuracy), 493.1253 (verification), 493.1255 (calibration), 493.1256 (QC), 493.1281 (comparisons) and 493.1291 (reporting).
Does this page list D-tag numbers?
No. D-tags come from CMS Appendix C; this map points to the 42 CFR 493 sections those tags are based on so you can read the rule itself.
What does Urine Drug Screen PT require?
Not listed in Subpart I: no required PT, but accuracy must be verified at least twice a year (42 CFR 493.1236(c)).
Sources
- 42 CFR 493.17, Test categorization
- 42 CFR 493.1236, Evaluation of proficiency testing performance
- 42 CFR 493.1253, Establishment and verification of performance specifications
- 42 CFR 493.1255, Calibration and calibration verification
- 42 CFR 493.1256, Control procedures
- 42 CFR 493.1281, Comparison of test results
- 42 CFR 493.1291, Test report
- CMS State Operations Manual, Appendix C (Laboratories)