The most-cited CLIA D-tags (CMS 2024 top 10)
CMS's Year 2024 data shows D5413 (test system and storage conditions) as the most-cited standard-level tag, at 818 laboratories (5.0%), and D2016 (successful proficiency testing participation) as the most-cited condition-level tag, at 738 laboratories (4.39%).

Top 10 standard-level D-tags (2024)
| D-tag | 42 CFR | Requirement (paraphrased) | Labs cited | % of labs |
|---|---|---|---|---|
| D5413 | §493.1252(b) | Conditions for reagent/specimen storage, test system operation and reporting defined, monitored, documented | 818 | 5.0% |
| D5209 | §493.1235 | Written policies to assess employee and consultant competency | 778 | 4.8% |
| D5401 | §493.1251(a) | Procedure manual available to and followed by personnel | 635 | 3.9% |
| D5403 | §493.1251(b) | Procedure manual content requirements | 609 | 3.7% |
| D5417 | §493.1252(d) | No expired, deteriorated or substandard reagents and supplies | 607 | 3.7% |
| D5217 | §493.1236(c)(1) | Twice-yearly accuracy verification for tests without PT | 548 | 3.4% |
| D5421 | §493.1253(b)(1) | Verify performance specifications of unmodified FDA-cleared tests | 451 | 2.8% |
| D6016 | §493.1407(e)(4)(i) | Director ensures PT samples tested as required | 451 | 2.8% |
| D5429 | §493.1254(a)(1) | Maintenance as defined by the manufacturer | 449 | 2.8% |
| D5805 | §493.1291(c) | Required test report content | 417 | 2.6% |
Top 10 condition-level D-tags (2024)
| D-tag | 42 CFR | Requirement (paraphrased) | Labs cited | % of labs |
|---|---|---|---|---|
| D2016 | §493.803 | Successful participation in PT | 738 | 4.39% |
| D6000 | §493.1403 | Moderate complexity laboratory director | 709 | 4.22% |
| D6076 | §493.1441 | High complexity laboratory director | 388 | 2.31% |
| D5400 | §493.1250 | Analytic systems | 340 | 2.02% |
| D2000 | §493.801 | PT enrollment and testing | 218 | 1.30% |
| D6033 | §493.1409 | Technical consultant | 189 | 1.13% |
| D6063 | §493.1421 | Testing personnel, moderate complexity | 182 | 1.08% |
| D6168 | §493.1487 | Testing personnel, high complexity | 172 | 1.02% |
| D6108 | §493.1447 | Technical supervisor | 110 | 0.65% |
| D5300 | §493.1240 | Preanalytic systems | 82 | 0.49% |
Source: CMS Division of Clinical Laboratory Improvement and Quality, Year 2024 Top 10 Standard-Level and Condition-Level Deficiencies. Requirement text is paraphrased.
What the pattern shows
- Documentation drives most standard-level tags. Seven of the ten are about defining, keeping and following records and procedures.
- Personnel dominates the condition list. Six of ten condition tags are Subpart M personnel conditions.
- PT appears at both levels. D2016 and D2000 are conditions; D6016 is the director's PT responsibility standard.
- New instruments are a risk point. D5421 is cited when verification is missing before patient testing.
Have a D-tag to answer?
Sound Lab Consultants maps each D-tag on your CMS-2567 to the regulation and Appendix C guidance and helps you build a correction that holds.
Related guides on soundlabconsult.com:
Frequently asked questions
What is the most common CLIA deficiency?
D5413 under 42 CFR 493.1252(b), cited at 818 laboratories in CMS's 2024 standard-level list.
What is the most common condition-level CLIA deficiency?
D2016, successful participation in proficiency testing (42 CFR 493.803), cited at 738 laboratories in 2024.
Where does CMS publish the top 10 list?
On the CMS CLIA website as the Top 10 deficiencies document from the Division of Clinical Laboratory Improvement and Quality.