Ionized Calcium: CLIA Requirements Mapped to 42 CFR 493
Ionized Calcium testing touches about eight CLIA standards, from test categorization (42 CFR 493.17) through QC (42 CFR 493.1256) and test reporting (42 CFR 493.1291). The table below maps each one to the Ionized Calcium records to keep ready.
Ionized Calcium requirement map
Use this map to find the regulation behind each Ionized Calcium requirement. Survey deficiency tags in CMS Appendix C are organized around these same sections.
| Requirement | 42 CFR 493 | What to keep on file for Ionized Calcium |
|---|---|---|
| Test categorization (waived, moderate, high) | 42 CFR 493.17 | FDA categorization printout for your Ionized Calcium test system |
| QC | 42 CFR 493.1256, 42 CFR 493.1267 | Daily Ionized Calcium QC records, Levey-Jennings charts, corrective action log, control lot ranges |
| Calibration verification | 42 CFR 493.1255 | Six-month Ionized Calcium calibration verification and reportable range data with director review |
| Accuracy verification | 42 CFR 493.1236 | Twice-yearly Ionized Calcium accuracy checks (external PT, split samples or reference material) with acceptance criteria |
| Comparison of results across methods or sites | 42 CFR 493.1281 | Twice-yearly comparison if Ionized Calcium runs on more than one instrument, method or site |
| Method verification | 42 CFR 493.1253 | Ionized Calcium verification study: accuracy, precision, reportable range, reference interval check, director sign-off |
| Test report and critical values | 42 CFR 493.1291 | Critical value list, call log with read-back, report format showing Ionized Calcium reference intervals |
| Personnel | Subpart M | Training and competency records for everyone who runs Ionized Calcium |
Proficiency testing
Not listed in Subpart I: no required PT, but accuracy must be verified at least twice a year (42 CFR 493.1236(c)).
Quick audit questions
- Can you show Ionized Calcium QC for every day patients were tested in the last month?
- When was the last Ionized Calcium calibration verification, and did it cover low, mid and high values?
- Is the Ionized Calcium verification study signed by the director and dated before the first patient result?
- Are Ionized Calcium critical value calls logged with read-back?
Related tests
Full guide: Ionized Calcium testing in the clinical lab: CLIA, PT and QC requirements on soundlabconsult.com, part of the Laboratory Test Compliance Library.
Have a D-tag to answer?
Sound Lab Consultants maps each D-tag on your CMS-2567 to the regulation and Appendix C guidance and helps you build a correction that holds.
Related guides on soundlabconsult.com:
Frequently asked questions
Which CLIA regulations apply to Ionized Calcium?
Mainly 42 CFR 493.17 (categorization), Subpart H/I or 493.1236 (PT and accuracy), 493.1253 (verification), 493.1255 (calibration), 493.1256 (QC), 493.1281 (comparisons) and 493.1291 (reporting).
Does this page list D-tag numbers?
No. D-tags come from CMS Appendix C; this map points to the 42 CFR 493 sections those tags are based on so you can read the rule itself.
What does Ionized Calcium PT require?
Not listed in Subpart I: no required PT, but accuracy must be verified at least twice a year (42 CFR 493.1236(c)).
Sources
- 42 CFR 493.17, Test categorization
- 42 CFR 493.1236, Evaluation of proficiency testing performance
- 42 CFR 493.1253, Establishment and verification of performance specifications
- 42 CFR 493.1255, Calibration and calibration verification
- 42 CFR 493.1256, Control procedures
- 42 CFR 493.1281, Comparison of test results
- 42 CFR 493.1291, Test report
- CMS State Operations Manual, Appendix C (Laboratories)