Blood Gases (pH, pO2, pCO2): CLIA Requirements Mapped to 42 CFR 493
Blood Gas testing touches about eight CLIA standards, from test categorization (42 CFR 493.17) through QC (42 CFR 493.1256) and test reporting (42 CFR 493.1291). The table below maps each one to the Blood Gas records to keep ready.
Blood Gas requirement map
Use this map to find the regulation behind each Blood Gas requirement. Survey deficiency tags in CMS Appendix C are organized around these same sections.
| Requirement | 42 CFR 493 | What to keep on file for Blood Gas |
|---|---|---|
| Test categorization (waived, moderate, high) | 42 CFR 493.17 | FDA categorization printout for your Blood Gas test system |
| QC | 42 CFR 493.1256, 42 CFR 493.1267 | Daily Blood Gas QC records, Levey-Jennings charts, corrective action log, control lot ranges |
| Calibration verification | 42 CFR 493.1255 | Six-month Blood Gas calibration verification and reportable range data with director review |
| Proficiency testing | 42 CFR 493.801, Subpart H | Blood Gas PT enrollment, attestation statements, graded reports, investigation of every miss |
| Comparison of results across methods or sites | 42 CFR 493.1281 | Twice-yearly comparison if Blood Gas runs on more than one instrument, method or site |
| Method verification | 42 CFR 493.1253 | Blood Gas verification study: accuracy, precision, reportable range, reference interval check, director sign-off |
| Test report and critical values | 42 CFR 493.1291 | Critical value list, call log with read-back, report format showing Blood Gas reference intervals |
| Personnel | Subpart M | Training and competency records for everyone who runs Blood Gas |
Proficiency testing
Blood Gases (pH, pO2, pCO2) is a regulated analyte (routine chemistry, 42 CFR 493.931). CLIA PT acceptance limit: pH: target value ±0.04; pCO2: target value ±8% or ±5 mm Hg (greater); pO2: target value ±15% or ±15 mmHg (greater).
Quick audit questions
- Can you show Blood Gas QC for every day patients were tested in the last month?
- When was the last Blood Gas calibration verification, and did it cover low, mid and high values?
- Is the Blood Gas verification study signed by the director and dated before the first patient result?
- Are Blood Gas critical value calls logged with read-back?
Related tests
Full guide: Blood Gases (pH, pO2, pCO2) testing in the clinical lab: CLIA, PT and QC requirements on soundlabconsult.com, part of the Laboratory Test Compliance Library.
Have a D-tag to answer?
Sound Lab Consultants maps each D-tag on your CMS-2567 to the regulation and Appendix C guidance and helps you build a correction that holds.
Related guides on soundlabconsult.com:
Frequently asked questions
Which CLIA regulations apply to Blood Gas?
Mainly 42 CFR 493.17 (categorization), Subpart H/I or 493.1236 (PT and accuracy), 493.1253 (verification), 493.1255 (calibration), 493.1256 (QC), 493.1281 (comparisons) and 493.1291 (reporting).
Does this page list D-tag numbers?
No. D-tags come from CMS Appendix C; this map points to the 42 CFR 493 sections those tags are based on so you can read the rule itself.
What does Blood Gas PT require?
Blood Gases (pH, pO2, pCO2) is a regulated analyte (routine chemistry, 42 CFR 493.931). CLIA PT acceptance limit: pH: target value ±0.04; pCO2: target value ±8% or ±5 mm Hg (greater); pO2: target value ±15% or ±15 mmHg (greater).
Sources
- 42 CFR 493.931, Routine chemistry
- 42 CFR 493.17, Test categorization
- 42 CFR 493.1236, Evaluation of proficiency testing performance
- 42 CFR 493.1253, Establishment and verification of performance specifications
- 42 CFR 493.1255, Calibration and calibration verification
- 42 CFR 493.1256, Control procedures
- 42 CFR 493.1281, Comparison of test results
- 42 CFR 493.1291, Test report
- CMS State Operations Manual, Appendix C (Laboratories)