Unexpected Antibody Detection (Antibody Screen): CLIA Requirements Mapped to 42 CFR 493
Antibody Screen testing touches about eight CLIA standards, from test categorization (42 CFR 493.17) through QC (42 CFR 493.1256) and test reporting (42 CFR 493.1291). The table below maps each one to the Antibody Screen records to keep ready.
Antibody Screen requirement map
Use this map to find the regulation behind each Antibody Screen requirement. Survey deficiency tags in CMS Appendix C are organized around these same sections.
| Requirement | 42 CFR 493 | What to keep on file for Antibody Screen |
|---|---|---|
| Test categorization (waived, moderate, high) | 42 CFR 493.17 | FDA categorization printout for your Antibody Screen test system |
| QC | 42 CFR 493.1256 | Daily Antibody Screen QC records, Levey-Jennings charts, corrective action log, control lot ranges |
| Proficiency testing | 42 CFR 493.801, Subpart H | Antibody Screen PT enrollment, attestation statements, graded reports, investigation of every miss |
| Method verification | 42 CFR 493.1253 | Antibody Screen verification study: accuracy, precision, reportable range, reference interval check, director sign-off |
| Comparison of results across methods or sites | 42 CFR 493.1281 | Twice-yearly comparison if Antibody Screen runs on more than one instrument, method or site |
| Test report and critical values | 42 CFR 493.1291 | Critical value list, call log with read-back, report format showing Antibody Screen reference intervals |
| Personnel | Subpart M | Training and competency records for everyone who runs Antibody Screen |
Proficiency testing
Unexpected Antibody Detection (Antibody Screen) is a regulated analyte (immunohematology, 42 CFR 493.959). CLIA PT acceptance limit: 100% accuracy.
Quick audit questions
- Can you show Antibody Screen QC for every day patients were tested in the last month?
- Is the Antibody Screen verification study signed by the director and dated before the first patient result?
- Are Antibody Screen critical value calls logged with read-back?
Related tests
Full guide: Unexpected Antibody Detection (Antibody Screen) testing in the clinical lab: CLIA, PT and QC requirements on soundlabconsult.com, part of the Laboratory Test Compliance Library.
Have a D-tag to answer?
Sound Lab Consultants maps each D-tag on your CMS-2567 to the regulation and Appendix C guidance and helps you build a correction that holds.
Related guides on soundlabconsult.com:
Frequently asked questions
Which CLIA regulations apply to Antibody Screen?
Mainly 42 CFR 493.17 (categorization), Subpart H/I or 493.1236 (PT and accuracy), 493.1253 (verification), 493.1255 (calibration), 493.1256 (QC), 493.1281 (comparisons) and 493.1291 (reporting).
Does this page list D-tag numbers?
No. D-tags come from CMS Appendix C; this map points to the 42 CFR 493 sections those tags are based on so you can read the rule itself.
What does Antibody Screen PT require?
Unexpected Antibody Detection (Antibody Screen) is a regulated analyte (immunohematology, 42 CFR 493.959). CLIA PT acceptance limit: 100% accuracy.
Sources
- 42 CFR 493.959, Immunohematology
- 42 CFR 493.17, Test categorization
- 42 CFR 493.1236, Evaluation of proficiency testing performance
- 42 CFR 493.1253, Establishment and verification of performance specifications
- 42 CFR 493.1255, Calibration and calibration verification
- 42 CFR 493.1256, Control procedures
- 42 CFR 493.1281, Comparison of test results
- 42 CFR 493.1291, Test report
- CMS State Operations Manual, Appendix C (Laboratories)