ABO Group: CLIA Requirements Mapped to 42 CFR 493
ABO Group testing touches about eight CLIA standards, from test categorization (42 CFR 493.17) through QC (42 CFR 493.1256) and test reporting (42 CFR 493.1291). The table below maps each one to the ABO Group records to keep ready.
ABO Group requirement map
Use this map to find the regulation behind each ABO Group requirement. Survey deficiency tags in CMS Appendix C are organized around these same sections.
| Requirement | 42 CFR 493 | What to keep on file for ABO Group |
|---|---|---|
| Test categorization (waived, moderate, high) | 42 CFR 493.17 | FDA categorization printout for your ABO Group test system |
| QC | 42 CFR 493.1256 | Daily ABO Group QC records, Levey-Jennings charts, corrective action log, control lot ranges |
| Proficiency testing | 42 CFR 493.801, Subpart H | ABO Group PT enrollment, attestation statements, graded reports, investigation of every miss |
| Method verification | 42 CFR 493.1253 | ABO Group verification study: accuracy, precision, reportable range, reference interval check, director sign-off |
| Comparison of results across methods or sites | 42 CFR 493.1281 | Twice-yearly comparison if ABO Group runs on more than one instrument, method or site |
| Test report and critical values | 42 CFR 493.1291 | Critical value list, call log with read-back, report format showing ABO Group reference intervals |
| Personnel | Subpart M | Training and competency records for everyone who runs ABO Group |
Proficiency testing
ABO Group is a regulated analyte (immunohematology, 42 CFR 493.959). CLIA PT acceptance limit: 100% accuracy.
Quick audit questions
- Can you show ABO Group QC for every day patients were tested in the last month?
- Is the ABO Group verification study signed by the director and dated before the first patient result?
- Are ABO Group critical value calls logged with read-back?
Related tests
Full guide: ABO Group testing in the clinical lab: CLIA, PT and QC requirements on soundlabconsult.com, part of the Laboratory Test Compliance Library.
Have a D-tag to answer?
Sound Lab Consultants maps each D-tag on your CMS-2567 to the regulation and Appendix C guidance and helps you build a correction that holds.
Related guides on soundlabconsult.com:
Frequently asked questions
Which CLIA regulations apply to ABO Group?
Mainly 42 CFR 493.17 (categorization), Subpart H/I or 493.1236 (PT and accuracy), 493.1253 (verification), 493.1255 (calibration), 493.1256 (QC), 493.1281 (comparisons) and 493.1291 (reporting).
Does this page list D-tag numbers?
No. D-tags come from CMS Appendix C; this map points to the 42 CFR 493 sections those tags are based on so you can read the rule itself.
What does ABO Group PT require?
ABO Group is a regulated analyte (immunohematology, 42 CFR 493.959). CLIA PT acceptance limit: 100% accuracy.
Sources
- 42 CFR 493.959, Immunohematology
- 42 CFR 493.17, Test categorization
- 42 CFR 493.1236, Evaluation of proficiency testing performance
- 42 CFR 493.1253, Establishment and verification of performance specifications
- 42 CFR 493.1255, Calibration and calibration verification
- 42 CFR 493.1256, Control procedures
- 42 CFR 493.1281, Comparison of test results
- 42 CFR 493.1291, Test report
- CMS State Operations Manual, Appendix C (Laboratories)