Radiometer method verification D-tags: citations after installation
If a laboratory reports patient results on a Radiometer analyzer without completed method verification, surveyors cite D5421 (or D5423 if specifications had to be established) under 42 CFR 493.1253. Calibration, calibration verification, QC, manufacturer maintenance and the procedure manual have their own D-tags. Sound Lab Consultants helps labs close those findings.

D-tags that apply when a Radiometer analyzer is cited
These tags come from CMS Appendix C and 42 CFR 493. They are not Radiometer-specific; surveyors use them whenever the underlying requirement is not met on any test system, including the FDA-listed Radiometer systems below.
| D-tag | Regulation | What surveyors look for when a new analyzer is in use |
|---|---|---|
| D5421 | 493.1253(b)(1) | Verification of performance specifications for an unmodified FDA-cleared or approved test system before patient results |
| D5423 | 493.1253(b)(2) | Establishment of performance specifications for a modified or non-FDA-cleared/approved method |
| D5425 | 493.1253(b)(3) | Determination of calibration and control procedures from verified or established specifications |
| D5427 | 493.1253(c) | Documentation of all verification or establishment activities |
| D5437 | 493.1255(a) | Calibration procedures following manufacturer instructions, materials and frequency, or equivalent laboratory criteria |
| D5439 | 493.1255(b) | Calibration verification at least every 6 months and when the rule otherwise requires it, including min, mid and max values |
| D5441 | 493.1256(a)–(b) | Control procedures that monitor the accuracy and precision of the complete analytic process |
| D5429 | 493.1254(a)(1) | Maintenance as defined by the manufacturer, at least as often as the manufacturer specifies |
| D5431 | 493.1254(a)(2) | Function checks as defined by the manufacturer, within established limits before patient testing |
| D5401 | 493.1251(a) | Written procedure manual available to and followed by testing personnel |
| D5413 | 493.1252(b) | Defined, monitored and documented conditions essential for reagent and specimen storage, test system operation and reporting |
| D2000 | 493.801 | Enrollment in an HHS-approved PT program for each specialty, subspecialty and regulated analyte |
Tag numbers and citations are from CMS State Operations Manual Appendix C (Laboratories). D5413 was the most-cited standard-level tag in CMS's Year 2024 list; D5421 (verification of unmodified FDA-cleared systems) was also among the more frequently cited analytic-system tags. See most-cited D-tags and the CMS Top 10 document.
FDA-listed Radiometer test systems these tags can attach to
In the FDA CLIA categorization download (clia_detail.zip, file dated November 23, 2026; 81,314 records with effective dates from 1993-07-26 through 2026-11-20; retrieved November 24, 2026, America/Phoenix), this manufacturer group accounts for 386 categorization records across 71 unique test system names: 13 waived, 368 moderate complexity, 5 high complexity. Those are counts of FDA records, not sales or installed-base figures. FDA records also list HemoCue test systems, a Radiometer company. Counts group the FDA-listed name variants shown on this page; they are not a claim of corporate identity for every historical name.
| FDA test system name | Complexity in the file | Analytes | Specialties |
|---|---|---|---|
| Radiometer ABL837 Flex Analyzer | Moderate (19) | 18 | General Chemistry, Hematology, Toxicology/TDM |
| Radiometer Medical ApS, ABL90 FLEX PLUS System | Moderate (19) | 14 | General Chemistry, Hematology, Toxicology/TDM |
| Radiometer Medical ApS, ABL90 FLEX | Moderate (18) | 17 | General Chemistry, Hematology, Toxicology/TDM |
| Radiometer Medical ApS, ABL90 FLEX PLUS | Moderate (17) | 16 | General Chemistry, Hematology, Toxicology/TDM |
| Radiometer ABL System 625 | Moderate (16) | 16 | General Chemistry, Hematology, Toxicology/TDM |
| Radiometer ABL827 Flex Analyzer | Moderate (16) | 15 | General Chemistry, Hematology, Toxicology/TDM |
| Radiometer ABL 700 Series | Moderate (15) | 15 | General Chemistry, Hematology, Toxicology/TDM |
| Radiometer ABL 620 | Moderate (14) | 14 | General Chemistry, Hematology, Toxicology/TDM |
| Radiometer ABL System 620 | Moderate (14) | 14 | General Chemistry, Hematology, Toxicology/TDM |
| Radiometer ABL817 Flex Analyzer | Moderate (13) | 12 | General Chemistry, Hematology |
| Radiometer ABL System 615 | Moderate (12) | 12 | General Chemistry, Hematology |
| Radiometer ABL 555 | Moderate (11) | 11 | General Chemistry, Hematology |
| Radiometer ABL800 FLEX with FLEXQ Module | Moderate (11) | 11 | General Chemistry, Hematology, Toxicology/TDM |
| Radiometer ABL 520 | Moderate (10) | 10 | General Chemistry, Hematology, Toxicology/TDM |
| Radiometer ABL System 605 | Moderate (10) | 10 | General Chemistry |
The table lists the 15 test system names with the most categorization records. The FDA file lists 71 unique test system names in this manufacturer group. Look up the exact analyzer and reagent in the FDA CLIA database before reporting patient results.
What good evidence looks like
- A director-approved verification or establishment plan and a report with the actual measurements (42 CFR 493.1253(c); D5427).
- Calibration records and a six-month calibration-verification packet with minimal, mid-point and maximum values (42 CFR 493.1255; D5437, D5439).
- Daily QC (or an approved equivalent / IQCP), with lot acceptability criteria (42 CFR 493.1256; D5441).
- Manufacturer maintenance and function-check logs that meet the instructions for use (42 CFR 493.1254; D5429, D5431).
- A procedure that testing personnel can follow (42 CFR 493.1251; D5401) and documented competency.
How Sound Lab Consultants helps
Sound Lab Consultants works with new and existing laboratories that are installing Radiometer analyzers, or adding assays to a system already on site. The firm is independent of the manufacturer. Typical work includes:
- Install planning: test menu, certificate type, space, utilities, LIS interface planning and a verification calendar that fits the go-live date.
- Method verification or establishment studies under 42 CFR 493.1253, including a written plan, data collection templates, calculations and a director-ready report.
- Procedure manuals for each assay, aligned with 42 CFR 493.1251 and the manufacturer's instructions for use.
- QC and, where used, IQCP design under 42 CFR 493.1256; calibration and calibration-verification schedules under 42 CFR 493.1255.
- Proficiency testing enrollment for the new specialty, subspecialty and analytes; competency assessments for testing personnel.
- Laboratory director and technical consultant oversight, including review of verification records before patient testing.
- CMS-116 and state-license test-menu updates, and inspection preparation for state agency or accreditation surveys.
- Startup of a new laboratory around a Radiometer core bench, or integration of a new analyzer into an existing quality system.
Work is available nationwide, remote or on site. No outcome is guaranteed; survey findings depend on the laboratory's own records and performance.
Cited on a Radiometer method?
The full cluster on soundlabconsult.com starts at the lab instrument manufacturer CLIA consulting hub, with one page per manufacturer. Book a free consultation with Sound Lab Consultants, an independent consulting firm.
Manufacturer names and trademarks are the property of their respective owners. Sound Lab Consultants is an independent consulting firm and is not affiliated with, authorized by, certified by or endorsed by Radiometer or any other manufacturer, the FDA, CMS, CAP, COLA, The Joint Commission or CDPH.
Have a D-tag to answer?
Sound Lab Consultants maps each D-tag on your CMS-2567 to the regulation and Appendix C guidance and helps you build a correction that holds.
Related guides on soundlabconsult.com:
Frequently asked questions
What is the usual tag if we started testing on a new Radiometer analyzer too soon?
D5421 when the system is an unmodified FDA-cleared or approved test system; D5423 when the laboratory had to establish performance specifications.
Where do these D-tag numbers come from?
CMS State Operations Manual Appendix C (Laboratories), which maps each tag to a 42 CFR 493 citation.
Does a Radiometer service visit replace verification?
No. Manufacturer installation or a service visit does not substitute for the laboratory's verification or establishment under 42 CFR 493.1253.
Can Sound Lab Consultants write the plan of correction?
Sound Lab Consultants helps laboratories analyze the citation, complete any missing studies and draft a plan of correction. Book a free consultation.
Sources
- CMS State Operations Manual, Appendix C (Laboratories)
- 42 CFR 493.1253, Establishment and verification of performance specifications
- 42 CFR 493.1255, Calibration and calibration verification procedures
- 42 CFR 493.1256, Control procedures
- 42 CFR 493.1254, Maintenance and function checks
- CMS CLIA Year 2024 Top 10 Standard- and Condition-Level Deficiencies
- FDA, Clinical Laboratory Improvement Amendments — Download Data (clia_detail.zip dated November 23, 2026; retrieved November 24, 2026)
- FDA CLIA test categorization database (search)