CLIA laboratory director deficiencies are cited under 42 CFR 493.1403 to 493.1407 (moderate complexity) and 493.1441 to 493.1445 (high complexity). Condition-level tags are D6000 and D6076. Because the director keeps ultimate responsibility even when duties are delegated, gaps in QC, PT, competency or written duties are often traced back to the director.

Last reviewed Nov 24, 2026 · By the Sound Lab Consultants regulatory team · Sources listed below

CLIA lab director deficiencies: D-tag and regulation map
CLIA lab director deficiencies: D-tag and regulation map

Why director citations matter

When a surveyor finds that the director is not qualified, or is not carrying out the director responsibilities, the finding can rise to a condition-level deficiency: D6000 for moderate complexity labs (42 CFR 493.1403) or D6076 for high complexity labs (42 CFR 493.1441). In CMS's CLIA Year 2024 Top 10 list, D6000 was cited at 709 laboratories (4.22%) and D6076 at 388 (2.31%), and the standard-level tag D6016 (director ensures PT samples are tested as required) at 451 (2.8%).

Director requirement to regulation map

Tag numbers for individual standards are listed in CMS State Operations Manual Appendix C. This map ties each director duty to its paragraph of 42 CFR 493 and the records that prove it.

Director requirement42 CFRD-tagWhat surveyors look for
Condition: moderate complexity director§493.1403D6000Director qualified under 493.1405 and performing 493.1407 duties
Condition: high complexity director§493.1441D6076Director qualified under 493.1443 and performing 493.1445 duties
Director ensures PT samples tested as required§493.1407(e)(4)(i)D6016PT attestations, testing records, same-as-patient handling
Onsite at least every 6 months, 4+ months apart; visits documented§493.1407(c) / 493.1445(c)see Appendix CDated visit log with evidence of director activities
No more than five laboratories directed§493.1407(d) / 493.1445(d)see Appendix CList of every CLIA certificate naming the director
Test systems cover pre-analytic, analytic, post-analytic phases§493.1407(e)(1) / 493.1445(e)(1)see Appendix CProcedures for collection, testing and reporting
Methods selected and verified§493.1407(e)(3)see Appendix CSigned method verification studies
PT enrollment, review and corrective action§493.1407(e)(4)see Appendix CEnrollment, graded reports, corrective action plans
QC and QA programs§493.1407(e)(5)-(6)see Appendix CQC review, QA plan and meeting minutes
Remedial action; results held until fixed§493.1407(e)(7)see Appendix CCorrective action log, held-result documentation
Sufficient qualified staff; training before testing§493.1407(e)(10)-(11)see Appendix CPersonnel files: diplomas, transcripts, training
Competency monitored in all phases§493.1407(e)(12)see Appendix CCompetency assessments at required intervals
Approved procedure manual§493.1407(e)(13)see Appendix CProcedures signed and dated by the director
Written duties for every consultant, supervisor and tester§493.1407(e)(14) / 493.1445(e)(15)see Appendix CDelegation letters, test authorization lists

New since December 28, 2024

If your director was cited

  1. Read the exact tag and the regulation paragraph cited on the CMS-2567.
  2. Decide whether it is a qualification problem (personnel file) or a performance problem (missing evidence of duties).
  3. Fix the system: delegation letters, a visit log, a review calendar for QC, PT and competency.
  4. Write the plan of correction with dates and a person responsible.

Need the whole picture on the director role?

The full CLIA laboratory director guide on soundlabconsult.com covers core responsibilities, retained responsibility when duties are delegated, qualifications by complexity, the five-lab limit, onsite visits and a director checklist. Book a free consultation with Sound Lab Consultants.

Have a D-tag to answer?

Sound Lab Consultants maps each D-tag on your CMS-2567 to the regulation and Appendix C guidance and helps you build a correction that holds.

Book a free consultation

Related guides on soundlabconsult.com:

Frequently asked questions

What D-tag is the CLIA lab director condition?

D6000 is the condition for the moderate complexity laboratory director (42 CFR 493.1403) and D6076 is the condition for the high complexity laboratory director (42 CFR 493.1441).

Can a lab be cited because of something the director delegated?

Yes. Under 42 CFR 493.1407(b) and 493.1445(b), a director who reapportions duties remains responsible for ensuring all duties are properly performed.

What evidence shows the director met the onsite requirement?

A dated visit log showing at least one visit every 6 months, at least 4 months apart, with evidence of director activities such as signed QC and PT review, procedure approvals and competency sign-offs.

How many labs can one director be cited for directing?

Each individual may direct no more than five laboratories (42 CFR 493.1407(d) and 493.1445(d)). CMS says the limit applies to nonwaived labs.