SARS-CoV-2 (COVID-19) Testing: CLIA Requirements Mapped to 42 CFR 493
COVID-19 testing touches about eight CLIA standards, from test categorization (42 CFR 493.17) through QC (42 CFR 493.1256) and test reporting (42 CFR 493.1291). The table below maps each one to the COVID-19 records to keep ready.
COVID-19 requirement map
Use this map to find the regulation behind each COVID-19 requirement. Survey deficiency tags in CMS Appendix C are organized around these same sections.
| Requirement | 42 CFR 493 | What to keep on file for COVID-19 |
|---|---|---|
| Test categorization (waived, moderate, high) | 42 CFR 493.17 | FDA categorization printout for your COVID-19 test system |
| QC | 42 CFR 493.1256 | Daily COVID-19 QC records, Levey-Jennings charts, corrective action log, control lot ranges |
| Proficiency testing | 42 CFR 493.801, Subpart H | COVID-19 PT enrollment, attestation statements, graded reports, investigation of every miss |
| Method verification | 42 CFR 493.1253 | COVID-19 verification study: accuracy, precision, reportable range, reference interval check, director sign-off |
| Comparison of results across methods or sites | 42 CFR 493.1281 | Twice-yearly comparison if COVID-19 runs on more than one instrument, method or site |
| Test report and critical values | 42 CFR 493.1291 | Critical value list, call log with read-back, report format showing COVID-19 reference intervals |
| Personnel | Subpart M | Training and competency records for everyone who runs COVID-19 |
Proficiency testing
No analyte-level limit; nonwaived labs enroll in virology PT for the services they perform (42 CFR 493.919).
Quick audit questions
- Can you show COVID-19 QC for every day patients were tested in the last month?
- Is the COVID-19 verification study signed by the director and dated before the first patient result?
- Are COVID-19 critical value calls logged with read-back?
Related tests
Full guide: SARS-CoV-2 (COVID-19) Testing testing in the clinical lab: CLIA, PT and QC requirements on soundlabconsult.com, part of the Laboratory Test Compliance Library.
Have a D-tag to answer?
Sound Lab Consultants maps each D-tag on your CMS-2567 to the regulation and Appendix C guidance and helps you build a correction that holds.
Related guides on soundlabconsult.com:
Frequently asked questions
Which CLIA regulations apply to COVID-19?
Mainly 42 CFR 493.17 (categorization), Subpart H/I or 493.1236 (PT and accuracy), 493.1253 (verification), 493.1255 (calibration), 493.1256 (QC), 493.1281 (comparisons) and 493.1291 (reporting).
Does this page list D-tag numbers?
No. D-tags come from CMS Appendix C; this map points to the 42 CFR 493 sections those tags are based on so you can read the rule itself.
What does COVID-19 PT require?
No analyte-level limit; nonwaived labs enroll in virology PT for the services they perform (42 CFR 493.919).
Sources
- 42 CFR 493.919, Virology
- 42 CFR 493.17, Test categorization
- 42 CFR 493.1236, Evaluation of proficiency testing performance
- 42 CFR 493.1253, Establishment and verification of performance specifications
- 42 CFR 493.1255, Calibration and calibration verification
- 42 CFR 493.1256, Control procedures
- 42 CFR 493.1281, Comparison of test results
- 42 CFR 493.1291, Test report
- CMS State Operations Manual, Appendix C (Laboratories)