NT-proBNP (proBNP): CLIA Requirements Mapped to 42 CFR 493
NT-proBNP testing touches about eight CLIA standards, from test categorization (42 CFR 493.17) through QC (42 CFR 493.1256) and test reporting (42 CFR 493.1291). The table below maps each one to the NT-proBNP records to keep ready.
NT-proBNP requirement map
Use this map to find the regulation behind each NT-proBNP requirement. Survey deficiency tags in CMS Appendix C are organized around these same sections.
| Requirement | 42 CFR 493 | What to keep on file for NT-proBNP |
|---|---|---|
| Test categorization (waived, moderate, high) | 42 CFR 493.17 | FDA categorization printout for your NT-proBNP test system |
| QC | 42 CFR 493.1256 | Daily NT-proBNP QC records, Levey-Jennings charts, corrective action log, control lot ranges |
| Calibration verification | 42 CFR 493.1255 | Six-month NT-proBNP calibration verification and reportable range data with director review |
| Proficiency testing | 42 CFR 493.801, Subpart H | NT-proBNP PT enrollment, attestation statements, graded reports, investigation of every miss |
| Comparison of results across methods or sites | 42 CFR 493.1281 | Twice-yearly comparison if NT-proBNP runs on more than one instrument, method or site |
| Method verification | 42 CFR 493.1253 | NT-proBNP verification study: accuracy, precision, reportable range, reference interval check, director sign-off |
| Test report and critical values | 42 CFR 493.1291 | Critical value list, call log with read-back, report format showing NT-proBNP reference intervals |
| Personnel | Subpart M | Training and competency records for everyone who runs NT-proBNP |
Proficiency testing
NT-proBNP (proBNP) is a regulated analyte (routine chemistry, 42 CFR 493.931). CLIA PT acceptance limit: target value ±30%.
Quick audit questions
- Can you show NT-proBNP QC for every day patients were tested in the last month?
- When was the last NT-proBNP calibration verification, and did it cover low, mid and high values?
- Is the NT-proBNP verification study signed by the director and dated before the first patient result?
- Are NT-proBNP critical value calls logged with read-back?
Related tests
Full guide: NT-proBNP (proBNP) testing in the clinical lab: CLIA, PT and QC requirements on soundlabconsult.com, part of the Laboratory Test Compliance Library.
Have a D-tag to answer?
Sound Lab Consultants maps each D-tag on your CMS-2567 to the regulation and Appendix C guidance and helps you build a correction that holds.
Related guides on soundlabconsult.com:
Frequently asked questions
Which CLIA regulations apply to NT-proBNP?
Mainly 42 CFR 493.17 (categorization), Subpart H/I or 493.1236 (PT and accuracy), 493.1253 (verification), 493.1255 (calibration), 493.1256 (QC), 493.1281 (comparisons) and 493.1291 (reporting).
Does this page list D-tag numbers?
No. D-tags come from CMS Appendix C; this map points to the 42 CFR 493 sections those tags are based on so you can read the rule itself.
What does NT-proBNP PT require?
NT-proBNP (proBNP) is a regulated analyte (routine chemistry, 42 CFR 493.931). CLIA PT acceptance limit: target value ±30%.
Sources
- 42 CFR 493.931, Routine chemistry
- 42 CFR 493.17, Test categorization
- 42 CFR 493.1236, Evaluation of proficiency testing performance
- 42 CFR 493.1253, Establishment and verification of performance specifications
- 42 CFR 493.1255, Calibration and calibration verification
- 42 CFR 493.1256, Control procedures
- 42 CFR 493.1281, Comparison of test results
- 42 CFR 493.1291, Test report